4. Transport, packaging and storage/5. Commissioning, operation

4. Transport, packaging and storage

4.1 Transport
Check the pressure sensor for any damage that may have been caused by transport.
Obvious damage must be reported immediately.
4.2 Packaging and storage
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change in installation site, sending for
repair).
Permissible conditions at the place of storage:
Storage temperature: -40 ... +70 °C [-40 ... +158 °F]
Humidity: 45 ... 75 % relative humidity (non-condensing)

5. Commissioning, operation

5.1 Checking the instrument

Prior to commissioning, the pressure sensor must be subjected to a visual inspection.
Leaking fluid is indicative of damage.
Only use the pressure sensor if it is in perfect condition with respect to safety.

5.2 Requirements for mounting point

The mounting point must meet the following conditions:
Protected from weather influences. Permanent exposure to UV light/sunlight can lead to a change in the colour of
the plastic parts and a clouding/yellowing of the status display. Therefore, a possible limitation of the visibility of the
status display cannot be excluded.
Under corrosive environmental conditions (such as salty, humid air), reductions in the gloss level of the metal
surfaces, or even corrosion on the instrument, may occur, which make readability of the product label more difficult.
Sealing faces are clean and undamaged.
Sufficient space for a safe electrical installation.
The instrument is vented to the atmosphere. Therefore, no coating or other covering may be applied which might
restrict the venting.
For information on tapped holes and welding sockets, see Technical information IN 00.14 at www.wika.com.
WIKA operating instructions, pressure sensor, model A-1200