P3 Top Class pin assignment
9
3
2
4
5
6
7
10
Information
If the measurement chain has no other connection to ground, earth or mains, its zero
potential can be connected to the transducer ground, by soldering an insulated, flexible
stranded wire from the cable shield to socket Y in the HK6S cable socket.
Information
To open the KH6S cable socket HK6S, loosen the clamping bolt (14 a.f.) from the sleeve
(15 a.f.) and pull the socket assembly together with the cable backward out of the sleeve.
The cable can then be pushed forward through a slit clamping sleeve and the spacer with
the seal, until the soldering tags and the cable shield are accessible.
Important
When removing the cable socket, please do not pull on the white socket assembly.
P3
ELECTRICAL CONNECTION
Measurement signal (+)
Measurement signal (-)
Br. excitation voltage (+)
Sense lead (+)
Br. excitation voltage (-)
Sense lead (-)
Sense lead
Excitation voltage
Sense lead
Excitation voltage
1
Lemo socket FFA.2E.310.CLAC65
8
View: pressure transducer
1
4
3
6
2
7
9
8
10
5
white
red
blue
green
black
gray
orange
yellow
violet
brown