When inductive consumers are switched off, this can cause voltage peaks and
can lead to adverse effects on electrical systems and may destroy the switching
contact.
Use commercially available standard RC combinations such as 0.1 F/
100 Ohm for inductive consumers.
The output signal is available via the relay output (voltage-free changeo-
ver contact). The alarm signal can be transmitted to a connected addi-
tional alarm unit (for example, ZAG 01).
If no alarm is present, the relay is de-energised; in case of an alarm, the
relay is energised and switches the changeover contact.
1. Permanently install the cable connected to the relay output.
2. Route this cable through the upper or bottom cable gland into the prod-
uct.
3. Connect the cable to the terminals designated with the relay output.
4. Verify safe separation of the relay cable.
- The insulation characteristics of the cable must at least comply with
IEC 60227 or IEC 60245.