Cleaning and disinfecting

Manual cleaning
EQUIPMENT
The method has been validated in Europe using:
– pH-neutral, enzymatic cleaning agent:
Bode Bomix
– Drinking water temperature of about 15 °C
– Application time: 10 minutes
PROCEDURE:
CAUTION
Risk of infection due to growth of bacteria
Inadequate disinfection encourages the growth of bacteria
and thus increases the risk of infection.
• Observe the specified mixing proportions and the spe-
cified treatment time.
• Ensure that all components are completely submerged in
the solution for the whole of the treatment time. There
must not be any air pockets or bubbles.
• Clean all individual parts with a solution prepared according
to the manufacturer's instructions.
For visible soiling, use a medium-soft brush (e.g. a tooth-
brush) which is used exclusively for this purpose.
If the recommended application period is exceeded signi-
ficantly, the plastic parts may take on the smell of the me-
dium used.
®
 plus (concentration: 0.1%)
– 30 –