Herunterladen Inhalt Inhalt Diese Seite drucken

Devicor mammotome elite Anleitung Seite 9

Biopsiesystem
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 25
Removing Probe from Holster
While pressing down on the Probe's Locking Tab, pull the Probe out of the Holster cavity (See Illustration 10). Dispose of
Probe in the appropriate container.
Illustration 10
CAUTION:
Probe is meant for SINGLE-USE ONLY. Dispose of Probe immediately after use.
CAUTION: Do not recap Probe tip.
Cleaning the Holster
After each use, manually clean the Holster thoroughly utilizing a pH neutral enzymatic detergent. Proceed with manual
cleaning utilizing an enzymatic detergent as follows:
1.
Prepare a pH neutral detergent or pH neutral enzymatic detergent according to the manufacturer's directions.
2.
Use a soft, clean cloth lightly moistened with the cleaning solution to manually clean all surfaces. In addition, use a
standard soft bristle toothbrush to clean areas containing crevices.
3.
Rinse thoroughly using a soft, clean cloth lightly moistened with lukewarm (27°C – 44°C) tap water.
4.
Dry with a clean, soft cloth.
5.
Inspect the device for visible contamination. If contamination is present, repeat steps 1-5 until no visible
contamination is present.
CAUTION:
Do not attempt to sterilize the Holster through autoclave, radiation, or plasma sterilization procedures; do
not spray with fluids or submerge in fluids; this may damage the instrument. If the Holster is cleaned improperly, the
warranty is void.
Disinfecting the Holster
A disinfection process must follow the cleaning procedure. The following chemical disinfectants are approved for use
with the Holster:
• 10% Bleach Solution (1 part bleach to 9 parts purified or deionized water)
Disinfectants should be prepared and used according to the manufacturer's recommendations for use, concentration
and contact time. The use of other disinfectants, other than those specified in these instructions, should be assessed
for equivalency before using. Technical data sheets are typically available through the manufacturer's web pages to
assist in this assessment. Any disinfection process, including tools and solutions, may influence the wear on a device or
equipment.
1.
Soak a clean cloth and a soft bristle toothbrush in the disinfectant solution. Squeeze the cloth thoroughly to
remove all excess solution. Tap toothbrush to remove excess solution.
2.
Thoroughly wipe down the Holster ensuring that a slightly wet surface is maintained for 5 minutes.
3.
Thoroughly scrub crevice areas with the soft bristle, standard toothbrush ensuring that a slightly wet surface is
maintained for 5 minutes.
4.
Sterile gauze or a clean, soft cloth moistened with purified water should be used to remove the disinfectant from
the device. Repeat 4 additional times to thoroughly remove residue.
5.
A clean soft bristle, standard toothbrush moistened with purified water should be used to remove the disinfectant
from the crevices of the device. Repeat 6 additional times to thoroughly remove residue.
6.
Dry the Holster with a clean, soft cloth and ensure the cloth is inserted into the crevices of the device.
9

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis