Application

10
Operating Instructions Sterisart
7
Application
Caution!
General danger with risk of injury
Do not operate the Sterisart
properly installed. The pump may only be operated with the tubing inserted
and only by one person at a time. Never pump any flammable, corrosive or
explosive fluids. Users should always wear tight-fitting clothing (no ties, etc.).
To be able to operate the pump, Sterisart
stalled beforehand. To do this, follow the steps given below:
Remove the sterility test systems from their respective packaging and place
the sterility test sample containers in the container for draining on the right
side of the pump.
— Now open the lever and the bow.
— Keep the tubing around the rotor of the pump (the arrow on the fixed
part surrounding the pump head indicates the subsequent direction of
rotation), so that no kinks occur and the tubing is not under tension (Fig. 3).
The tubing is attached by simply inserting it (with it lying on the ground)
and fastening the lever (Fig. 4). In this way, the tubing is automatically cor-
rectly positioned.
Important Note!
When threading the tubing, make sure that it is not kinked or twisted.
Fig. 3
— To start the pumping process, the tubing must be fully inserted and the
lever must be engaged. A safety function ensures that the pump does
not start if the lever or the bow are open.
Universal Pump 16419
®
Universal pump until the tubing has been
®
sterility test systems must be in-
®