W&H Lisa log64 Bedienungsanleitung Seite 12

Digitaler zyklusspeicher
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 28
Led
Memory
Status
Off
Green
steady
Red/Green
blinking
Red
blinking
Off
Red steady
Orange
blinking
Off
Orange
Orange
Orange
Orange
ACOUSTIC SIGNALING
Signaling
Description
4 beeps
CABLE DISCONNECTED
every 15
The parallel cable is disconnected. If a cycle is launched the data will not be saved. Check the
minutes
connection between the sterilizer and Lisa log 64.
MEMORY NEARLY FULL
3 beeps at the
It is possible to save data, but the remaining memory capacity is less than 50 cycles. Transfer
beginning of
the data onto your PC and insert a new one. It is also possible to erase data from the previous
the cycle
one and re-use it.
MEMORY FULL, MISSING or FAILED
Impossible to save data.
Even if the memory card is full, Lisa log 64 is able to memorize a complete sterilization cycle.
Insert an empty card only at the end of the sterilization cycle.
3 beeps each
time the data
CAUTION:
is received
If other cycles are started the new received data could be lost. Transfer the data onto
your PC and insert a new one. It is also possible to erase data from the previous one
and re-use it.
MEANING
St-by
Stand-by status or power supply is missing.
OK – the data can be saved on the memory card
OK – the data can be saved, but the remaining memory capacity is less
than 50 cycles. Transfer the data onto your PC and insert a new one. It is
also possible to erase data from the previous one and re-use it.
Impossible to save data – cable disconnected.
Off
Impossible to save data– see Memory LED status.
Warning: the last file saved does not contain a complete data cycle.
Check the parallel cable and perform a reprint (see page 7).
Note:
The signalling continues until a complete file is saved; during a
cycle this signalling does not mean that there is an abnormality.
With at least one LED on, the recorder is OK.
Off
With all the LEDs off, no power supply
Yellow
OK – Stand-by status
Power supply problems. Check the cable connections (Connect and
disconnect).
If the problem remains, Lisa log 64 or the power supply may be damaged
– call for service.
Off
Note: When Lisa log 64 and a printer (via a special adaptor) are used
at the same time and the printer is turned on, the signalling may
remain on even if the power supply is not connected to the Lisa
log 64.
Yellow
Device not functioning (try to turn it off and then back on)
9

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis