Signal description

LED
Three Led's are available for a quick view of the recorder condition:
-
Memory – Status LED of the memory card,
-
Status - Status LED of the recorder module,
-
St-by - Stand-by status LED
The presence of acoustic and/or red signalling means that there are abnormalities and/or possible loss of data.
Other combinations of signalling are not associated with anomalies, but provide an overview of the operating status
of the device.
Led
Memory
Status
Off
Red blinking
+
MMC missing
Green steady
Green
blinking
Off
Red/Green
blinking
Red blinking
+
MMC plugged
in
Red steady
8
MultiMediaCard
plugged in
MEANING
St-by
Memory card missing, Stand-by status or power supply missing
Memory card missing during the cycle - To save the data, plug in the card
as soon as the sterilization cycle ends.
WARNING:
Do not remove or plug in the memory card during a
sterilization cycle.
Lisa log 64 is able to memorize a complete sterilization cycle
without memory card plugged in. To save the sterilization cycle
on the memory card the card must be inserted after the
sterilization cycle.
It is possible to start a sterilization cycle, and/or transfer the data onto the
computer
Lisa log 64 is receiving data. Do not unplug the memory card.
The data will be saved onto the memory card at the end of the cycle,
indicated with a faster blinking of the LED.
Off
Note: If the cycle is not ended and no data is being received for more
than 45 minutes (Sterilizer Off) the file will be saved as
incomplete.
The diagnostic of the memory card is in process (from 2 to 40 seconds)
The memory card is defective or not formatted
The memory card is full. It is impossible to save other cycle data!
WARNING:
Do not remove or plug in the memory card during a
sterilization cycle.
Even if the memory card is full, Lisa log 64 is able to memorize
a complete sterilization cycle. Insert an empty card only at the
end of the sterilization cycle.