Herunterladen Inhalt Inhalt Diese Seite drucken

Recognized Buffers; Maintenance; Clean The Instrument; Sensor Cleaning - HACH LANGE sensION+ PH 3 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4. Operation

4.7. Recognized buffers

Buffer solutions
+
The sensION
PH 3
values as a function of the temperature).
Technical buffers (DIN 19267)
ºC
ºF
0
32
2.01
10
50
2.01
20
68
2.00
25
77
2.00
30
86
2.00
40
104
2.00
50
122
2.00
60
140
2.00
70
158
2.01
80
176
2.01
90
194
2.01
sensION+ PH3 HACH_LANGE 0710.indd ANG:9
sensION+ PH3 HACH_LANGE 0710.indd ANG:9
All manuals and user guides at all-guides.com
has in memory the following data (pH
pH
4.01
7.12
9.52
10.30
4.00
7.06
9.38
10.17
245
4.00
7.02
9.26
10.06
228
4.01
7.00
9.21 10.01
220
4.01
6.99
9.16
9.96
212
4.03
6.97
9.06
9.88
195
4.06
6.97
8.99
9.82
178
4.10
6.98
8.93
9.76
160
4.16
7.00
8.88
--
142
4.22
7.04
8.83
--
4.30
7.09
8.79
--

6. Maintenance

6.1. Clean the instrument

Important note: Never use cleaning such as turpentine, acetone
or similar products to clean the instrument including the display
and accessories.
Only clean the housing and accessories using a soft, damp cloth.
Mild soap solution may also be used. Dry the cleaned parts care-
fully with a soft cotton cloth.
mV
--
CAUTION
When using chemicals or solvents, comply with the instruc-
tions of the producer and all local safety regulations.

6.2. Sensor cleaning

See the manual of the sensor in use.
--
--
9
8/7/10 16:39:54
8/7/10 16:39:54

Werbung

Inhaltsverzeichnis
loading