TECHNICAL SPECIFICATIONS (CONT)
Cooling time of the device when it has been stored at 45°C
for it to be ready for use at ambient temperature (20°C)
dimensions
actiTENS
Weight of
actiTENS
charger unit dimensions
actiCHARGE
charger unit weight
actiCHARGE
IP rating
actiTENS
charger unit IP rating
actiTENS
Waveform
Pulse width
Frequency
Treatment time
Electrode
electrode storage conditions
actiTENS
adhesive pad storage conditions
actiTENS
Composition
actiTENS
Composition
actiTENS
Composition of the sensitive skin electrodes
Composition of the lumbar electrode
Composition of the electrodes
adhesive pad
textile accessories
10 minutes minimum
108 mm x 53.5 mm x 17 mm
~ 65 g
133.8 mm x 79 mm x 34 mm
~ 115g
IP22: protection against solid bodies greater
than 22.5 mm, protection against dripping water
up to 15° from the vertical
IP21: protection against solid bodies
greater than 12.5 mm, protection against
vertically dripping water
Compensated asymmetric two-phase waves
50-400 µs ± 5 µs
1-120Hz ± 10%
10-720 min adjustable with the mobile app
Any electrode in which the current density
exceeds 2 mA/cm
requires special
2
attention; stimulation should never be
painful.
From 5°C to 27°C
From 0°C to 40°C
Medical grade silicone
Tissu principal : Polyamide 58%,
Polyester 30%, Elasthanne 12%
Etiquette : Polyamide 90%, Polyester 10%
Boucle et velcro : Polyamide
Hydrogel biocompatible conducteur, partie
conductrice : carbone/ argent, matériau de
base isolant : tricot, connecteur snap : acier
inoxydable
Hydrogel biocompatible conducteur, partie
conductrice : carbone vynile, matériau de
base isolant : non tissé PET 25 microns,
Velcro : velour 3165, connecteur snap :
acier inoxydable 316
Biocompatible hydrogel conductor,
conducting part: carbon + PE; insulating
base material: non-woven PET fibre, snap
connector: stainless steel
91