P8 FREQUENCY MODULATION
Technical specifications
Mode of action
Frequency (Hz)
Fréquense (Hz)
Fréquense (Hz)
Séquence 1
Séquence 1
Sequence 1
P9 MUSCLE STIMULATION
Intensité (mA)
Intensité (mA)
Technical specifications
Travail
Travail
Mode of action
1
sequence (duration: 7.5 s):
st
Rising frequency: 2 Hz to 80 Hz
Falling pulse width 200 µs to 100 µs
2
sequence (duration: 7.5 s):
nd
Falling frequency: 80 Hz to 2 Hz
Rising pulse width 100 µs to 200 µs
These two sequences are alternated every 7.5 s.
Default duration 30 min
Combined action
Inhibition of pain signal
Stimulation of endorphin secretion for a general
analgesic effect
Séquence 2
Séquence 2
Sequence 2
Frequency 50 Hz
Pulse width 250 µs
1
sequence (2 s):
st
Increasing intensity: 0 mA to the set intensity
2
sequence (5 s):
nd
constant intensity (set intensity)
Pause
Pause
3
sequence (1 s):
rd
Decreasing intensity set intensity to 0 mA
4
sequence (12 s):
th
No pulse
The sequences are alternated for the duration of the
programme
Default duration 30 min
Muscle strengthening
Fréquense (Hz)
Fréquense (Hz)
Temps
Temps
Time
(secondes)
(secondes)
(seconds)
Fréquense (Hz)
Fréquense (Hz)
Temps
Temps
(secondes)
(secondes)
Séquence 1
Séquence 1
Séquence 1
Séquence 1
85