25. TECHNICAL DATA SHEET
Intended operator
Channels
Programmes
Battery
output voltage
actiTENS
Essential performance of the
Power supply input voltage
Input and output voltage of the
charger unit
actiCHARGE
Conditions of use of the
charger unit
actiCHARGE
storage conditions
actiTENS
charger unit storage
actiCHARGE
conditions
Warm-up time of the device when it has
been stored at -10°C for it to be ready for
use at ambient temperature (20°C)
90
TECHNICAL SPECIFICATIONS
actiTENS
and the
actiTENS
Device user
2 independent channels
13 programmes including:
Gate control : 80-100Hz, pulse width 150-200 µs
Endorphinic: 2Hz, pulse width 250 µs7
Mixed programmes
EMS programme (electrical muscle stimulation)
Current delivered: from 1 mA to 60 mA ± 10% in
steps of 0.5 mA
Li-ion
max 60 mA ± 10% (1000 Ω) / max 60V
Does not deliver a current 60 mA ± 10% or a voltage
> 60 V
100-240 V AC 0.2-0.1 A
5 V DC – 1 A
From 10°C to 40°C with relative humidity from 15%
to 93%
Atmospheric pressure from 700 hPa to 1060 hPa
From -10°C to 45°C with max. relative humidity of
93%
From -20°C to 60°C
5 minutes minimum
SBM1AL001 – Notice_202105_v8