Table 1. Troubleshooting issues
Problem
The transmitter is con‐
nected to the charger
and no lights come on.
During charging, the
flashing green light on
the charger turns off and
a longer flashing red
light appears on the
charger.
When charging, the
flashing green light on
the charger turns off and
there is a series of quick
flashing red lights on the
charger for two seconds
at a time.
During charging, a mix
of quick and long flash‐
ing red lights appear on
the charger.
Likely Cause(s)
The transmitter connec‐
tor pins are damaged or
corroded.
The charger battery has
no power or no battery is
inserted.
The charger battery is
low on power.
The transmitter is low on
power.
The charger and the
transmitter are low on
power.
Resolution
1. Check the transmitter connector pins for
damage or corrosion. For more informa‐
tion about the connector pins, see
Inspecting the transmitter connector
pins, page 22. If the pins are damaged or
corroded, contact a local Medtronic sup‐
port representative. It may be time to
replace the transmitter.
2. If there is no damage to the connector
pins, replace the battery in the charger.
For instructions on replacing the charger
battery, see Installing a battery in the
charger, page 17.
Replace the battery in the charger. For instruc‐
tions on replacing the charger battery, see
Installing a battery in the charger, page 17.
1. Charge the transmitter continuously for
one hour. If flashing does not stop, pro‐
ceed to step 2.
2. Charge the transmitter continuously for
eight hours. If flashing does not stop, call
a local Medtronic support representa‐
tive. It may be time to replace the trans‐
mitter.
1. Replace the battery in the charger. For
instructions on replacing the charger
battery, see Installing a battery in the
charger, page 17.
2. Charge the transmitter continuously for
one hour. If the quick flashing red lights
do not stop, proceed to step 3.
3. Charge the transmitter continuously for
eight hours. If flashing does not stop, call
a local Medtronic support representa‐
34