Herunterladen Inhalt Inhalt Diese Seite drucken

Medtronic GUARDIAN 4 Bedienungsanleitung Seite 22

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 43
CAUTION: Only use the green colored tester (MMT-7736L) with the transmitter. Pockets on
the transmitter are visible when connected to the tester. Do not use any other test plug. Other
test plugs are not intended for use with the transmitter and will damage the transmitter and
the tester.
Inspecting the transmitter connector pins
This image is an example of how the connector pins should look for the transmitter.
Figure 4. Transmitter components
Look inside the connector opening of the transmitter to make sure that the connector pins are not
damaged or corroded. If the connector pins are damaged or corroded, the transmitter cannot
communicate with the charger or pump. Contact a local Medtronic support representative. It may be
time to replace your transmitter.
Look for moisture inside the connector opening. If any moisture is present, allow the transmitter to dry
for at least one hour. Moisture inside the connector opening could cause the transmitter to not work
properly and could cause corrosion and damage over time.
Connecting the tester for testing or cleaning
Before proceeding, have the MiniMed 780G System User Guide available.
green
tester
connector opening
connector pins
housing
22
pockets

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis