Herunterladen Inhalt Inhalt Diese Seite drucken

Medtronic GUARDIAN 4 Bedienungsanleitung Seite 15

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 43
• Do not twist the tester or sensor while attached to the transmitter. Twisting the tester or sensor will
damage the transmitter.
• Do not allow the tester to come in contact with any liquid when not connected to the transmitter.
A wet tester can damage the transmitter.
• Do not allow the transmitter to come in contact with any liquid when not connected to a sensor
or to the tester. Moisture will damage the transmitter and a wet transmitter can damage the sensor.
• Do not clean the O-rings on the tester with any substances. Cleaning the O-rings can damage the
tester.
Figure 2. O-rings
IEC 60601-1-2:2014, 4th Edition; Special EMC Precautions for Medical Electrical
Equipment
1. Special Precautions regarding Electromagnetic Compatibility (EMC): This body worn device is
intended to be operated within a reasonable residential, domestic, public or work environment
where common levels of radiated "E" (V/m) or "H" fields (A/m) exist, such as cellular phones,
Wi-Fi™*, Bluetooth
This device generates, uses, and can radiate radio frequency energy and, if not installed and used
in accordance with the provided instructions, may cause harmful interference to radio
communications.
2. Portable and mobile RF communications equipment can affect medical electrical equipment.
If you encounter RF interference from a mobile or stationary RF transmitter, move away from the
RF transmitter that is causing the interference.
3. Be careful when using the transmitter closer than 30 cm (12 in) to portable radio frequency (RF)
equipment or electrical equipment. If the transmitter must be used next to portable RF
wireless technology, electric can openers, microwave and induction ovens.
®
O-rings
15

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis