Installation
8
Installation
8.1
Installation conditions
8.1.1
Orientation
Not permissible!
5
Permitted orientations
The sensor must be installed with an inclination angle of 10 ˚ to 170 ˚ in an assembly,
support or a corresponding process connection. Recommended angle: 45° in order to avoid the
formation of air bubbles.
Inclination angles other than those mentioned are not permitted. Do not install the sensor
overhead.
Observe the instructions for installing sensors in the Operating Instructions for the
assembly used.
8.1.2
Mounting location
• Choose a mounting location that can be easily accessed at a later stage.
• Ensure that upright posts and fittings are fully secured and vibration-free.
• Select an installation location which produces a typical oxygen concentration for the
individual application.
22
Permissible angle of installation
10°
Oxymax COS22/22D
10° Not permissible!