4. Packaging

Has everything been supplied?
Check the scope of supply:
Completely assembled pressure transmitters.
With flush version (F-21) including pre-assembled sealings and protection cap.
Inspect the pressure transmitter for possible damage during transportation. Should there be any obvious damage,
inform the transport company and WIKA without delay.
Keep the packaging, as it offers optimal protection during transportation (e.g. changing installation location,
shipment for repair).
Ensure that the pressure connection thread and the connection contacts will not be damaged.
In order to protect the diaphragm, the pressure connection of the instrument F-21 is provided with a special protection cap.
Remove this protection cap only just before installing the pressure transmitter in order to prevent any damage to the
diaphragm or the thread.
Keep the protection cap of the pressure connection thread and the diaphragm for later storage or transport.
Mount the protection cap when removing and transporting the instrument.
5. Starting, operation
Required tools: wrench (flats 27), screw driver
Diaphragm test for your safety
It is necessary that before starting the pressure transmitter you test the diaphragm visually, as this is a safety-relevant
component.
WARNING!
Pay attention to any liquid leaking out, for this points to a diaphragm damage.
Check the diaphragm visually for any damage (F-21).
Use the pressure transmitter only if the diaphragm is undamaged.
Use the pressure transmitter only if it is in a faultless condition as far as the safety-relevant features
are concerned.