Operation
Resistance measurement
Set the selector switch to measurement range 'Ω'. Insert the black test lead
into the 'COM' socket and the red test lead into the 'V Ω A' socket. Using the
test probes, touch the test points of the test object. The multimeter
automatically searches for the most suitable measurement range. Read
measurement value on the display.
Resistance
Range
200 Ω
2k Ω
20 kΩ
200 kΩ
2 MΩ
20 MΩ
Measurement voltage: 0.25 V.
Over-range protection: 600 V AC / DC < 30 s.
Diode
Set the selector switch to measurement range 'Ω'. Insert the black test lead
into the 'COM' socket and the red test lead into the right-hand socket. Use
the 'M' button to set the device to
points of the test object. Red test lead = anode, Black test lead = cathode.
The forward voltage drop is displayed.
Diode test
Range
Forward biasing current: approx. 0.6 mA, backward voltage: approx. 1.5 V.
Over-range protection: 600 V AC / DC < 30 s.
32
Resolution
0.1 Ω over-range protection:
250 V AC / DC
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
. Using the test probes, touch the test
Resolution
0.001 V
Accuracy
± 1% + 5 digits
± 1 % + 5 digits
± 1.8 % + 5 digits
Accuracy
Displays the forward voltage
drop
Testboy
®
TB-3000