Envision™
3. Apply exposed adhesive directly to area of interest on patient.
NOTE: Envision may be repositioned on patient, as desired.
WARNING
Ensure no air bubbles are present prior to use. Air left between cover and transducer lens may cause poor
n
image quality.
4. Inspect cover to ensure there are no holes or tears.
5. Use suitable hydration source, such as liquid saline or a saline wipe, to fully saturate Envision.
NOTE: Envision may be hydrated multiple times. Rehydrate as necessary.
6. Ensure Envision is fully hydrated prior to use.
4