5 Cleaning, care and disinfection

The Petermann Stand-up Aid Pegasus is made from rust-proof components and therefore doesn't require any
special maintenance. A visual inspection for defects is required before each use. The device may not be used if there
are any noticeable defects. Therefore, please contact your responsible medical supply store or call our service
number. In accordance with the hoist standard (DIN EN ISO 10535:2006), a periodic inspection must be performed
at least once a year; i.e. visual inspection of the load-bearing parts and the lifting mechanism (function test). Please
contact your medical supply store in this regard.
For sanitary reasons, please clean the device on a regular basis. You can use common liquid household cleaner.
After cleaning, all moving parts such as the threaded rod, gears, etc. can be coated with an oil-free lubricant
(Vaseline) using a brush.
If you do not wish to perform the inspection, cleaning and disinfection of your Stand-up Aid Pegasus yourself, you
can send your device to our shop.
The device can be cleaned and/or disinfected as follows before being used again:
Cleaning / disinfection of the device as follows is recommended:
Place the Stand-up Aid Pegasus without the crank and accessories in a suitable tub with lid. Fill with suitable
cleaning solution (Refer to the detergent specifications for mixing ratio and residence time).
Final cleaning with pressure washer. Addition of a disinfectant leads to simultaneous disinfection of the
device.
Rinsing and drying
Visual inspection for cleanliness and integrity (e.g. material properties, corrosion)
Function test of the frame / scissor and spring mechanism
Acoustic check: unusual noises during raising or lowering?
Inspect the seat and backrest
Repeat the cleaning and disinfection process, if necessary
Test run and verification of functionality under maximum load
Inspection of the accessories as well as suction cup, crank for proper functionality
General repair, replacing of accessories, if necessary
Final function test of the device
In the event of wear or damage to the device which cannot be repaired, the Stand-up Aid Pegasus must
be taken out of service!
Please let us know of any damage or defects on the device which are not attributable to normal wear and
tear, so that we can implement corrective measures if necessary.

6 Technical Data

Item no.
Material
PM-4222
Drive mechanism:
Seat:
Colour
white
 - high grade stainless
steel and spring steal
19/38
Load-bearing capacity
135 kg