Contraindications

Do not use without a suitable pressure redistribution surface and package of care.
Do not use with patients with unstable spinal cord injuries or who are undergoing traction.
Do not use with patients with equilibrium disturbances.
Advice for laypersons
Any change in skin condition that causes concern, should be reported to a healthcare professional.

Warnings and Precautions

Individual patient assessments should be made prior to use and appropriate pressure redistributing
devices put in place. Please note the following:
Unusual body shapes and postures should be assessed prior to use.
Caution is advised when using with patients with underlying neurological disorders which may
lead to increased anxiety.
Other medical devices should be assessed for suitability.
Ensure the Toto platform is fully deflated before attempting to leave the bed.
Engage therapy pause mode during hygiene procedures and when over-bed tables are being
used to avoid the risk of spillages.
When elevating a profiling bed back rest, raise the knee break first.
Ensure the surface of the bed frame is clear from debris before placing the Toto platform, as this
can damage the coating of the cover.
The system should not be used without a mattress.
Do not strap the mattress to the bed frame as this will prevent Toto from operating.
If mattress overlays are being used, these must be securely fastened to the mattress, not the
Toto platform.
Toto can be used with positioning systems, bed levers and linen cradles, if their support frame
fits horizontally across the bed. These must be positioned under the Toto platform.
Evacuation hammocks must be placed between the Toto platform and the mattress.
If using foot board protectors, assess for compatibility.
Do not place the control unit directly underneath the bed frame if placing on the floor, to avoid
lowering the bed onto the control unit.
Use only with the supplied AC power lead and with correct fuse fitted. Should there be a need
to replace the fuse, ensure only the correct fuse rating is used (3A).