1.5
Skin Preparation
Excessive hair, dirty skin, dry skin, or oily skin can
impact the quality of the ECG tracing. Clean the
patient's skin with water and soap and remove excess
hair as needed to achieve the highest quality tracing.
Do not use the DUO over wound areas or areas of
broken skin. Do not use an alcohol based skin cleaner
as this dries out the patient's skin and increases
resistance. Dry the skin vigorously to increase capillary
blood flow to the tissues.
ECG gels or saline solutions can be used on the
electrodes to improve signal quality.
1.6
EMC Compliance
FCC Intentional Radiator Certification
Contains FCC ID: 2ANB3-E5
Contains IC: 23063-E5
This equipment contains an intentional radiator
approved by the FCC under the FCC ID numbers
shown above. This device complies with Part 15 of
the FCC rules. Operation is subject to the following
two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any
interference received, including interference that may
cause undesirable operation.
NO MODIFICATION: Modifications to this device
shall not be made without the written consent of Eko
Devices, Inc. Unauthorized modifications may void
the authority granted under Federal Communications
Commission rules permitting the operation of this
device.
EMC Compliance Europe
This equipment complies with the EMC requirements
of the IEC 60601-1-2.
1.7
Cleaning
To ensure a proper ECG and PCG recording, clean the
electrodes with an alcohol-based cleaning solution
prior to use.
Cleaning and Disinfecting Procedure
The device and earpieces should be cleaned between
each patient use. All cleaning instructions pertaining
to stethoscopes and ECG units in general apply.
Under normal conditions it is unnecessary to separate
the device from the earpieces for cleaning. All external
parts of the hardware can be cleaned with 70%
isopropyl alcohol wipes.
NOTE: DO NOT immerse the device in any liquid or
subject it to any high-pressure/autoclave sterilization
processes.
P.