Installation
20
Instructions for Use Sartocon
5.3.2
Integrity Test
Each cassette is fully tested at Sartorius. It is therefore recommended that
no other integrity test be carried out on-site.
Cassette configuration
Hydrosart
Sartocon
®
®
with 0.1 to 0.42 cm² surface area
PESU Sartocon
Slice Self Contained
®
with 0.1 to 0.42 cm² surface area
Hydrosart
and PESU Sartocube
®
with 0.6 to 3.5 m² surface area
Hydrosart
and PESU Sartocube
®
with 7 m² surface area
Hydrosart
and PESU Sartocube
®
with 14 m² surface area
5.4
Information about Complaints
If you would like to send cassettes to Sartorius Stedim Biotech for review,
please ensure the following:
− that the customer service is contacted, who will provide you with a return
authorization form.
− that the cassettes have been completely cleaned;
− that the cassettes have been disinfected;
− that the cassettes are provided with the following information:
− the medium that was filtered using the cassettes;
− any errors that were observed;
− the operating parameters in use at the time of filtration (pressure
settings, operating time, process cycles, cleaning conditions,
operating temperature);
− cleaning method used;
− disinfecting agent used.
Slice Self Contained | Sartocube
®
Slice Self Contained
Self Contained
®
Self Contained
®
Self Contained
®
Self Contained
®
Integrity test limit at
1 bar test pressure
≤ 5 ml/min
≤ 15 ml/min
≤ 50 ml/min
≤ 75 ml/min
≤ 100 ml/min