3. Safety Notes

3. Safety Notes

3.1 General safety notes

This safety class 1 instrument was built and
tested in accordance with:
EN 61010-1:2001,
IEC 61010-1:2001,
Safety requirements for electrical equipment for
measurement, control and laboratory use.
Caution!
To maintain the unit in its original condition
and to ensure safe operation, the user must
follow the instructions and warnings con-
tained in this User Manual.
Caution!
The instruments and accessories described
in this operating manual have been tested for
safety and potential hazards.
The responsible Leica affi liate or the main
plant in Wetzlar must be consulted whenever
the instrument is altered, modifi ed or used in
conjunction with non-Leica components that
are outside of the scope of this manual.
Unauthorized alterations to the instrument or
noncompliant use shall void all rights to any
warranty claims and product liability
8

3.2 Electrical safety

General specifi cations
For indoor use only.
Supply voltage:
Frequency:
Power input:
Fuse (power supply):
Ambient temperature:
Relative humidity:
Overvoltage category:
Pollution degree:
Caution!
The power plug may only be plugged into an
outlet equipped with a grounding contact.
Do not interfere with the grounding function
by using an extension cord without a ground
wire. Any interruption of the ground wire in-
side or outside of the instrument, or release
of the ground wire connection, can cause the
instrument to become hazardous. Intentional
ground interruption is not permitted!
Type label Leica FS M
100-240 VAC
50-60 Hz
Imax 0.5 A
2 x 2,5 A H 250 V
15-35°C
max. 80% up to 30°C
II
2