Responsibility of manufacturer in case of
repair
The manufacturer shall only be responsible for the effects
on the safety, reliability and performance of the device, if
Ø the installation, extension, readjustment, alteration
or repair is carried out by technically experienced
and trained staff and
Ø the electrical installation of the respective room
complies with the requirements of the IEC guidelines
Ø the device is used as specified in the Instructions for
Use.
Ø Where any form of repair, i.e. change of printed
circuit board etc. has been carried out on an opened
device, the final test, described in section 12 must be
carried out. Only original AESCULAP spare parts may
be used for any repairs.
AESCULAP offers respective training courses, after which
trained members of staff are authorized to repair the
devices covered during the training course.
1

Measuring and test equipment

1.
Oscilloscope:
2.
Universal
digital
multimeter
3.
Safety tester
50MHz / two-channel, where
possible, digital storage with
respective key buttons
For measuring voltage and
resistance
i.e. Rigel 233, GERB GM-200,
Bender Safety Tester Mod. 601,
etc.
2

Opening and closing the device

7 7 7 7
Warning:
Ø Disconnect from mains prior to opening the
device as appliance sockets, cables or other
contact parts inside the device may still contain
hazardous mains voltages.
2.1

Opening the device

C
A)
Remove the 4 screws on the base
B)
Fold back the housing cover
Caution: the earthing cable of the cover may
not be excessively tensioned!
C)
Disconnect the protective conductor from the
cover
2.2

Closing the device

Note:
Prior to replacing the cover:
• ensure the correct connection between the cover and
the earthing cable
• inspect the device as specified in section 3. (visual
inspection)
Ø Pull bottom edges of device cover slightly apart and
position over device.
Ø Ensure that no leads are caught in between the
cover and the device.
Ø Replace the 4 screws on the base.