Setting up the instrument
3.2
Charging the battery
Fig. 23: Removing the batteries
Fig. 24: Battery charger
32
NOTE!
If the Wöhler CFM 600 has not been used for a
long time, the battery must necessarily be
charged before measuring.
Attention!
The battery must be removed from the device
before charging.
Remove the batteries from the Wöhler CFM
•
600 as follows:
Rotate the inner black knurl of the locking
•
screw ( Marked with an arrow in the adjacent
figure) in counterclockwise direction, until it
unscrews.
•
Remove the battery pack from the threaded
rod.
Charge the battery with the battery charger.
•
During charging, the charging status is indi-
cated by the colors of the LEDs.
Yellow: No battery pack is connected.
Orange: Fast charging.
Green-Yellow: Battery nearly completely
charged.
Green: Battery completely charged.
Orange-Green: Error
Attention!
If you do not want to use the Wöhler CFM 600
for a longer period, charge the battery and store
it in the carrying case (disconnected form the
device).