WARNING
Do not flush dry gas continuously for unnecessarily long
periods through the soda lime in the anesthesia system!
Otherwise the soda lime will dehydrate. If the humidity falls
below a minimum set point, undesirable reactions can
occur, independent of the type of lime and the volatile
anesthetic agent being used:
–
reducted CO
absorption,
2
–
increased heat generation in the CO
therefore increased breathing gas temperature,
–
CO formation,
–
absorption and/or breakdown of the volatile anesthetic
agent.
The above mentioned reactions may endanger the patient,
e.g.:
–
CO poisoning
–
insufficient depth of anesthesia
–
burns of the airways
Replacing
The soda lime must be replaced at the latest if one of the following
conditions applies:
–
The inspiratory CO
concentration has reached approx.
2
5 mmHg.
–
Approximately 2/3 of the soda lime have changed to violet.
The color indicator may slowly regenerate and the soda lime
may revert to white, nevertheless its absorption ability is spent!
–
The soda lime has been at the anesthesia workstation for
1 week.
The soda lime must be replaced, even if the soda lime has not
been used!
Principle of operation
Drägersorb 800 Plus binds the CO
in an exothermic reaction. The hydroxides are converted into
carbonates and water.
Characteristics / Technical data
Ambient conditions
During operation
Temperature
10 °C to 40 °C (50 °F to 104 °F)
Atmospheric
500 to 1200 hPa (7.3 psi to 17.4 psi)
pressure
Relative humidity
20 to 100 %
During storage and transport
Temperature
–20 °C to +50 °C (-4 °F to 122 °F)
Atmospheric
500 to 1200 hPa (7.3 psi to 17.4 psi)
pressure
Relative humidity
20 to 100 %
Composition
calcium hydroxide, natrium hydroxide, ethyl
violet, water
Form / diameter
half sphere / approx. 4 mm
Color change
white to violet
Instructions for use Drägersorb 800 Plus
absorber and
2
contained in the breathing gas
2
Classification
Class IIa
according to EC
Directive 93/42/EEC
Annex IX
UMDNS code
17-509
Universal Medical
Device
Nomenclature
System –
Nomenclature for
medical devices
Storage and transport
–
Store and transport the soda lime in its original packing only.
Once opened, resealable packs must be used within 1 month.
A briefly perceptible odor after opening the pack is normal and
insignificant.
–
Protect the pack from heat and direct sunlight.
–
Never use after the expiration date.
Maximum storage duration
The maximum storage duration is 4 years. See information on the
product label of the canister. The canister must be disposed of
after the expiration date has expired.
Reprocessing and disposal
CAUTION
The spent disposable CO
absorber cannot be prepared for
2
reuse.
–
At the end of life, dispose of the product in accordance with
applicable regulations.
–
It is recommended that the possibility of disposing of soda lime
in household waste be checked with local waste disposal
authorities.
–
In Europe, the applicable waste material code for soda lime is
EWC 160303/180106.
List of accessories
Drägersorb 800 Plus
Canister wrench
Safety data sheet
Drägersorb 800 Plus safety data sheet available on request.