How to carry out an adjustment:
 Confirm menu setting „noLin" by
Ensure that there are no objects on the weighing plate.
Wait for stability display, then press
The currently set adjustment weight will be displayed.
To change by using the navigation buttons (see chap. 2.1.1)
select the desired setting, the active digit is flashing.
Acknowledge with
Carefully place adjusting weight in the centre of the weighing
plate. Wait for stability display, then press
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will return
into weighing mode automatically.
An adjusting error or incorrect adjusting weight will be
indicated by the error message; repeat adjustment
procedure.
24
.
.
.
.
SFB-BA-e-1630