9. VISUALISATION OF A/D CONVERTER POINTS

- Upon start-up, press simultaneously the ZERO and % keys during the countdown.
- The instrument shows the A/D converter points relative to the weight on the scale.
- Press M+ to exit the step.

10. GRAVITY OF USE ZONE

Upon start-up, press simultaneously the F and % keys during the countdown: through this operation one access the
acceleration value setting of instrument use zone.
Manual entry of the g value: Set the value by using the F key to clear the value, the TARE key to increase the value of the
digit, the % key to go on the previous digit and the +/- key to go on the following digit; one must modify all the 6 digits of
the gravitational acceleration. Confirm the entered value with ZERO or press M+ to exit the step.

11. SEALING ACCESS TO THE BALANCE SETTINGS

For legal for trade applications, the balance must be sealed to prevent access to the
To seal the balance, press the
message "LEGAL" appears an instant), then cover the Lock switch access hole. A cover seal and paper seal may be
installed as shown.
To regain access to the locked balance settings, break the seal and press the recessed Lock switch
momentarily during power up (the message "HIRES" appears an instant).

12. ERROR MESSAGES

ERROR MESSAGES DESCRIPTION
- - - - - -
Over range
Err 4
Zero
Error
Err 5
Keyboard Error.
Err 6
Input signal
greater than
3mV/V

RECYCLING INSTRUCTION

The crossed-out wheeled bin on the product means that at the product end of life, it must be taken to
separate collection or to the reseller when a new equivalent type of equipment is purchased. The
adequate differentiated refuse collection in having the product recycled, helps to avoid possible
negative effects on the environment and health and supports the recycling of the materials of which
the equipment is made. The unlawful disposal of the product by the user will entail fines foreseen by
the current regulations.
recessed
Lock switch at the bottom of the balance, momentarily during power up (the
RESOLUTION
Remove weight from the scale.
Setting
Upon start-up or when the ZERO key is pressed, the weight on the scale
exceeds the percentage programmed on the max. capacity.
Remove the excess weight and retry.
The keyboard could be damaged.
Indicates that the A/D converter value is greater than the maximum value:
- Remove the weight from the scale if it's exceeding.
- The load cell or the electronics could be damaged.
metrological