Herunterladen Inhalt Inhalt Diese Seite drucken

Calibration; Procedure With Defined Sample Weights; Procedure With Free Sample Weight - Scale House AWM Serie Bedienungshandbuch

Multifunktions-/zählwaage
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

oF
AUTO SWITCH-OFF
By pressing TARE one sets the minutes of not being used for the auto switch-off (see section 6.10):
0
(disabled)
3
(3 minutes of not being used)
5
(5 minutes of not being used)
15
(15 minutes of not being used)
30
(30 minutes of not being used)

8. CALIBRATION

Premise: The unit of measure of the calibration is fixed at kilogram ("kg").
It's possible to calibrate the scale by using two defined sample weights or by manually entering a sample weight (can be
useful if one does not have defined sample weights).
IN CASE THE ZONE OF USE IS DIFFERENT FROM THE CALIBRATION ZONE ONE SHOULD:
1) Carry out the calibration as following described.
2) Set the gravity acceleration value for the ZONE OF USER, see section 10.

PROCEDURE WITH DEFINED SAMPLE WEIGHTS

- Upon start-up, press simultaneously an instant the +/- and % keys during the countdown.
- The display shows the acceleration value set: enter the gravitational acceleration value of the calibration zone if different
from the use zone, by using the F key to clear the value, the TARE key to increase the value of the digit, the % key to go
on the previous digit and the +/- key to go on the following digit. One must modify all the 6 digits of the gravitational
acceleration.
- Confirm the entered value with ZERO.
- The display shows "LOAD 0";
- Empty the platter, after Stable indicator on, press ZERO key to sure;
- The display shows "LOAD 1";
- Put the first sample weight, defined in the table below, on the scale and press ZERO when the stability indicator is on;
- The display shows "LOAD 2";
- Put the second sample weight, defined in the table below, on the scale and press ZERO when the stability indicator is on;
- The display shows "LOAD 3";
- Put the third sample weight, defined in the table below, on the scale and press ZERO when the stability indicator is on;
- The indicator returns to weighing.
MODEL
LOAD 1
LOAD 2
LOAD 3

PROCEDURE WITH FREE SAMPLE WEIGHT

- Upon start-up, press simultaneously an instant the TARE and % keys during the countdown.
- The display shows the acceleration value set: enter the gravitational acceleration value of the calibration zone if
different from the use zone, by using the F key to clear the value, the TARE key to increase the value of the digit,
the % key to go on the previous digit and the +/- key to go on the following digit. One must modify all the 6 digits of
the gravitational acceleration.
- Confirm the entered value with ZERO.
- The display shows "UNLOAD";
- Empty the platter, after Stable indicator on, press ZERO key to sure;
- Set the value of the mass weight by using the F key to clear the value, the TARE key to increase the value of the digit,
the % key to go on the previous digit and the +/- key to go on the following digit.
- Confirm the entered value with ZERO; the display shows "LOAD";
- Put the mass on platter, after stable indicator on, press ZERO key to sure;
- The indicator returns to weighing.
CALIBRATION WEIGHTS
AWM3
1 kg
2 kg
3 kg
AWM6
2 kg
4 kg
6 kg
14
AWM15
5 kg
10 kg
15 kg
AWM30
10 kg
15 kg
30 kg

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Awm3Awm6Awm15Awm30

Inhaltsverzeichnis