5. STORAGE

After the disinfection cycle, it is imperative that the unit is stored appropriately. This
involves using the Clean Storage Cover (900PT603) or an alternative method which
will keep the outlet elbow free from contamination while being stored between
patients.
USING THE CLEAN STORAGE COVER (900PT603)
A IRVO
900PT603
Clean Storage Cover
Susie Q
Name:
3rd Feb
Date:
21
Cycle #:
FPH hospital
Belongs to:
Do not use if
seal is broken
ALTERNATIVE STORAGE METHOD
If the Clean Storage Cover is not used, the alternative storage method is to leave the
disinfection tube and filter attached to the AIRVO 2. Store the unit (with disinfection
tube and filter attached) in an area suitable for medical devices, which is clean, dry
and dust-free.
Remove the disinfection tube and filter from the
AIRVO 2. To remove the disinfection tube, always push
the sleeve up to avoid tube damage.
Wrap the AIRVO 2 in the Clean Storage Cover, such
that the identification label sits prominently above the
display and buttons of the AIRVO 2.
To record the previous disinfection process:
Fill in the details on the identification label.
Seal the Clean Storage Cover with the adhesive tabs.
After the unit is sealed:
Store the unit and disinfection kit in an area suitable
for medical devices, which is clean, dry and dust-free.