2. FIT THE BLUE FILTER
Fit the blue filter to the right hand chamber port as
shown.
Push it firmly into position.
3. START THE DISINFECTION CYCLE
Connect the power cable to the mains/utility power
socket and start the disinfection cycle by holding the
On/Off button for 3 seconds until the unit beeps.
WARNINGS
• Disconnect any oxygen supply from the unit before starting
the disinfection cycle.
4. WARM-UP AND CALIBRATION
The unit will first go through a series of calibration
checks for approximately 2 minutes, during which a
warm-up symbol will be displayed on screen.
5. DURING DISINFECTION
After a successful calibration, the disinfection cycle
begins and runs for 55 minutes.
A – 7
Press the Mode button at any
time to view a graph of the
current disinfection cycle.
The graph shows:
Time (mins)
Temperature (°C)
If no button has been pushed
for 1 minute, a screensaver
will be launched.