Herunterladen Diese Seite drucken
Vorschau ausblenden Andere Handbücher für Airvo 2:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 15

3. CLEANING

WARNINGS
• The products above have been tested by Fisher & Paykel Healthcare. If none of these are
available, other cleaning agents may be used as long as they are pH-neutral, non-abrasive,
non-toxic and non-corrosive. Do not use any cleaning agents incompatible with polycarbonate
plastic (including, but not limited to, ammonia, ammonium hydroxide, caustic soda, iodine,
methanol, methylated spirits, turpentine and alkaline bleaches such as sodium hypochlorite).
• Always remove any residue with a clean, damp, lint-free cloth.
This cleaning step aims to remove any contamination from
the outlet elbow (between the right-hand chamber port
and the heated breathing tube connection port), before the
disinfection cycle is carried out. It is important to carry out
this cleaning step first to ensure that the disinfection cycle
will be effective. In addition, it is important to clean the other
outside surfaces of the unit.
Do not clean the left-hand chamber port, as this can damage
the non-return valve located at this port.
THE FOLLOWING EQUIPMENT IS REQUIRED:
Cleaning Sponge-Stick (900PT602)
• Available from Fisher & Paykel Healthcare.
Mild detergent
70% alcohol solution OR 70% alcohol wipes
Clean, disposable, lint-free cloths
A – 4

Quicklinks ausblenden:

Werbung

loading