Precooling of the rotor

For quick precooling of the unloaded rotor and the accessories, it is a good idea to do a centrifugation run with
the continuous operation setting and a speed of about 20% of the maximum rotor speed.
The precooling is done automatically with the PREC (PRECOOLING) program.
Precooling is not possible if program linkages are being used.
Press the
key. The LED in the
count starts from 00:00.
Press the
key to end the centrifugation run. Run- down is carried out with the selected run- down
OPEN / STOP
parameters. The run-down parameter is displayed e.g.
Once the rotor has come to a standstill, the LED in the
displayed. The right-hand LED in the
starts blinking and continues blinking until the lid is opened.
During the centrifugation run, the speed of the rotor, or the resulting RCF value, the sample temperature and the
remaining time or expired time are displayed.
31

Heating (only on centrifuges with heating/cooling option)

During operation of the centrifuge the centrifuge chamber can be heated as required to the preselected temperature.
The heating is switched off when the rotor is stationary.
Risk of burning! The surface temperature of the heating element in the centrifugal chamber of the centrifuge
can reach 500°C / 932°F. Do not touch the heating element.
Plastic suspension can only be used in temperatures up to a maximum of 40°C / 104°F.
32

Relative centrifugal force (RCF)

The relative centrifugal force (RCF) is given as a multiple of the acceleration of gravity (g). It is a unit-free value and
serves to compare the separation and sedimentation performance.
These values are calculated using the formula below:
RCF
RCF = relative centrifugal force
RPM = rotational speed (revolutions per minute)
r = centrifugal radius in mm = distance from the centre of the turning axis to the bottom of the centrifuge.
The relative centrifugal force (RCF) stands in relation to the revolutions per minute and the centrifugal
radius.
48/134
key blinks until the rotor has been imported, it is subsequently lit. The time
key also switches off, the left LED in the
OPEN / STOP
2
RPM
=
×
×
r
1,118
1000
For more on the centrifugal radius see the chapter " Anhang/Appendix,
Rotoren und Zubehör/Rotors and accessories".
9. The right-hand LED in the
key switches off and
OPEN
RCF
=
RPM
×
r
1,118
key is lit.
OPEN / STOP
OEFFNEN is
key
OPEN / STOP
×
1000