Centrifugation with time selection

Set the time or call up a program with a pre-set time (see Chapter "Entering centrifugation parameters" or
"Calling up programs").
Press the
START
When the time has expired or if the centrifugation run is interrupted by pressing the
place with the selected run-down parameters. When the rotor comes to a stop, the
display until the lid is opened.
During the centrifugation run, the speed of the rotor or the RCF value resulting from this, the sample temperature (for
centrifuges with cooling only) and the remaining time will be displayed.
15.2

Continuous running

Select the ---:-- symbol or call up a continuous running program (see Chapter "Entering centrifugation
parameters" or "Calling up programs").
Press the
START
from 00:00.
Press the
STOP
parameters. When the rotor comes to a stop, the
During the centrifugation run, the speed of the rotor or the RCF value resulting from this, the sample temperature (for
centrifuges with cooling only) and the time run will be displayed.
16

Changing settings during the centrifugation run.

The running time, the speed, the relative centrifugal force (RCF/RZB), the run-up and run-down parameters as well
as the temperature (for centrifuges with cooling only) can be changed while the centrifuge is running.
The parameters can only be changed one at a time and in succession.
Select the desired parameter and change the value using the control knob
centrifugation parameters").
Press the
START
buffer"). The original program will not be overwritten.
17
Integral RCF (∫RCF)
The integral RCF (∫RCF) is a measure for the sedimentation effect (∫ n
centrifugation runs. To interrogate the integral RCF, press and hold the
18

Displaying the maximum speed of the rotor

Press the
n
key repeatedly until the parameter RPM is displayed and the input field is shown inversely.
Press the
key once more and hold it pressed; the maximum speed of the rotor (n-max-Rotor) will be displayed.
n
19

Displaying the maximum RCF of the rotor

Press the
RCF
Press the
RCF
displayed.
20

Emergency stop

Press the
STOP
In the case of an emergency stop, run-down takes place with braking step R9 (shortest run-down time). If braking
step R0 was selected, the run-down time will be longer than with braking step R9 for technical reasons.
38/102
key. The rotation indicator
key. The rotation indicator
key to stop the centrifugation run. Run-down will take place with the selected run-down
key. The modified setting will be stored in program location "----" (see Chapter "Automatic
key repeatedly until the parameter RCF/RZB is displayed and the input field is shown inversely.
key once more and hold it pressed; the maximum RCF of the rotor (RCF-max-Rotor) will be
key twice. The STOP symbol will flash in the display.
will appear as long as the rotor is turning.
will appear as long as the rotor is turning. The time count starts
symbol will flash in the display until the lid is opened.
2
dt). This numerical value is used to compare
key.
RCF
key, run-down will take
STOP
symbol will flash in the
(see Chapter "Entering

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

5005-08a5005-085005-885005-98

Inhaltsverzeichnis