Herunterladen Inhalt Inhalt Diese Seite drucken

Enter Centrifugation Parameters; Running Time; Speed (Rpm) - Hettich ROTINA 380 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für ROTINA 380:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Call up programs and program linkages; parameter RCL.
PROG
Programs: Program locations 1 to 99 and PREC. program linkages: Program locations A to Z.
Save programs and program linkages; parameter STO.
99 programs can be saved (for centrifuges without refrigeration: program places 1 to 99, for centrifuges
with refrigeration: program places 1 98 and PREC). The program place PREC (PRECOOLING) is
reserved for the precooling program. Program place 0 is used for temporary storage for centrifuge data
from the last centrifugation run. No programs can be saved at this program place.
It is possible to save 25 program linkages (program locations A to Z; there is no program location J). A
program linkage can consist of 20 programs.
Link programs; parameter EDIT.
Start "Machine Menu" (keep the key held down for 8 seconds) and select the parameters from this menu.
Start rotor precooling (only for centrifuges with refrigeration). The LED in the key is lit during the
centrifugation run, as long as the rotor is revolving.
Precooling of the rotor takes place automatically with the PREC (PRECOOLING) program.
Start centrifugation run. The LED in the key is lit during the centrifugation run, as long as the rotor is
START
revolving.
Short-time centrifugation.
The centrifugation run occurs while the key is kept pressed. The LED in the key is lit during the
centrifugation run, as long as the rotor is revolving.
Saving entries and changes.
End centrifugation run.
STOP
OPEN
The rotor decelerates with the preset run-down parameters. The right-hand LED in the button lights up
until the rotor is stationary. Once the rotor is stationary the left-hand LED flashes in the button. Pressing
the button twice triggers the EMERGENCY STOP.
Unlock the lid.
The left-hand LED in the button goes out.
Leave parameter input and the "Machine Menu".
15

Enter centrifugation parameters

If no key is pressed for 8 seconds long after the selection or during the input of parameters, the previous
values will be shown in the display. The input of parameter then has to be executed again.
When you enter several parameters, the
settings for the last parameter.
If parameters are changed, the program place number is displayed in parentheses (). This means that the
centrifugation data in the display no longer corresponds to the centrifugation data from the program place that
has been saved.
You can no discontinue entering parameters at any time by pressing the key
adjustments are not saved.
15.1

Running time

In order to set continual running, the minutes, seconds and hours have to be set to zero.
Continual running is represented in the display by the following symbol, "∞".
Press the
key. The parameters t/hms is displayed. The minutes (m) are shown in parentheses < >, and
TIME
can be changed.
Use the adjusting knob
Press the
key. The seconds (s) are shown in parentheses < > and can be changed.
TIME
Use the adjusting knob
Press the
key. The hours (h) are shown in parentheses < > and can be changed.
TIME
Use the adjusting knob
Press the
key, to have the settings applied to the display.
START
15.2

Speed (RPM)

Press the
key. The parameters RPM is displayed.
RPM
Use the adjusting knob
Press the
key, to have the settings applied to the display.
START
key does not have to be pressed until you have made the
START
to set the value you want.
to set the value you want.
to set the value you want.
to set the value you want.
EN
. In this case the
OPEN / STOP