Centrifugation

When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous substances or
objects may be within the safety margin of 300 mm around the centrifuge.
If the permissible weight difference within the rotor loading has been exceeded, the drive shuts down during
the start-up, and IMBALANCE is displayed.
If the speed in the selected program is higher than the maximum speed of the rotor (Nmax), it will not be
possible to start a centrifugation run. The message N > ROTOR MAX will be displayed (see Chapter "Faults").
If N > ROTOR MAX in Prog : e.g. 5, Runtime 00:00 in Prog : e.g. 5 or Empty Program is shown during
program linkage, no centrifuge operation can be started (see the chapter on " Faults ").
A centrifugation run can be stopped at any time by pushing the key
All parameters can be selected and altered during the centrifugation run (see Chapter "Enter centrifugation
parameters").
It is possible to switch over between RPM and RCF display at any time using the
Switching is not possible if program linkages are being used. When working with the RCF display, it is
necessary to enter the centrifuging radius.
If
OPEN
OEFFNEN is displayed, a further operation of the centrifuge is only possible after opening the lid
once.
If the rotor has been changed, no centrifugation run takes place and this display is shown, e.g.
Rotor 4 Nmax= 4500 R=184 mm (see chapter "Rotor Identification").
Operation errors and malfunctions will be shown (see Chapter "Faults").
Switch on the mains supply switch. Switch position Ι.
Load the rotor and close the centrifuge lid.
17.1

Centrifugation with preselected time

Set a run time, call up a program with preset time or a program linkage (see the chapters on "Enter centrifugation
parameters", "Calling up programs" or "Program linkage").
Press the
key. The LED in the
START
Once the time has run up or if the centrifugation run is canceled by pressing the
takes place with the selected run-down parameter. The run-down parameter is displayed e.g.
hand LED in the
OPEN / STOP
the right LED in the
OPEN / STOP
During the centrifugation run, the speed of the rotor or the RCF value resulting from this, the sample temperature (for
centrifuges with cooling only) and the remaining time will be displayed.
17.2

Continuous operation

Set the minutes, seconds and hours to "0" or start a continual run program (see chapter "Enter centrifugation
parameter" or " Calling up programs").
Press the
key. The LED in the
START
The time count starts from 00:00.
Press the
key to end the centrifugation run. Run- down is carried out with the selected run- down
OPEN / STOP
parameters. The run-down parameter is displayed e.g.
Once the rotor has come to a standstill, the LED in the
switch off and the left LED in the
During the centrifugation run, the speed of the rotor or the RCF value resulting from this, the sample temperature (for
centrifuges with cooling only) and the time run will be displayed.
key blinks until the rotor has been imported, it is subsequently lit.
START
key is lit. Once the rotor has come to a standstill, the LED in the
key switch off and the left LED in the
key blinks until the rotor has been imported, it is subsequently lit.
START
9. The right-hand LED in the
START
key lights up.
OPEN / STOP
.
OPEN / STOP
OPEN / STOP
key lights up.
OPEN / STOP
OPEN / STOP
key and the right LED in the
EN
and
keys.
RPM
RCF
key, run-down
9. The right-
key and
START
key is lit.
key
OPEN / STOP
43/134