Unpacking the centrifuge

Lift the carton upward and remove the padding.
Do not lift by the front panel.
Observe the weight of the centrifuge, refer to chapter "Technical specifications".
Lift the centrifuge on both sides with an appropriate number of helpers and place it on the laboratory table.
8

Initial operation

According to the laboratory instrument standards EN / IEC 61010-2-020 an emergency switch to separate power
supply in the event of a failure must be installed in the building electrical system.
This switch has to be placed remote from the centrifuge, preferred outside of the room in which the centrifuge is
installed or near by the exit of this room.
Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety
margin of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous
substances or objects may be within the safety margin of 300 mm around the centrifuge.
Do not place any object in front of the ventiduct.
Keep a ventilation area of 300 mm around the ventiduct.
Centrifuge with PE connector:
If required, connect the PE connector at the back of the device with an additional medical potential equalization
system.
Centrifuge with RS232 interface:
Connect the RS232 centrifuge interface with an RS232 connecting cable (not included in the scope of delivery) to
the PC.
Check whether the mains voltage tallies with the statement on the type plate.
Connect the centrifuge with the connection cable to a standard mains socket. For connection ratings refer to
Chapter "Technical specifications".
Switch on the mains switch (switch position "Ι"). The LEDs in the keys are blinking.
The following displays appear one after the other:
1. The centrifuge model
2. The program version
3. The rotorcode (rotor), the maximum rotor speed (Nmax) and the preset centrifuging radius (R) of the last rotor
that was recognized by rotor recognition.
4.
OPEN
OEFFNEN
Open the lid.
The centrifuge data of the last used program or program 1 will be displayed.
Remove the transport safety device (see instruction sheet on "Moving the equipment safely").
9

Interface (only for centrifuge with interface)

As an option, the device can be equipped with an RS232 interface.
The RS232 interface is labeled with the
The centrifuge can be controlled and data queried via this interface.
The LED in the
key lights up during data communication.
PROG
10

Opening and closing the lid

10.1

Opening the lid

The lid can only be opened when the centrifuge is switched on and the rotor is at rest. If it cannot be
opened under these circumstances, see the section on "Emergency release".
Press the button
OPEN / STOP
extinguishes.