18.4. Inspect the device
Test for leakage:
 Consumption of compressed air is stopped.
 Pressurize the air tank by releasing some air via a connected device.
 With the compressor off, wait at least 10 minutes before checking the pressure level.
 If the pressure has dropped, use soapy water to find the leaky joint. Start with the joints outside
the case, then test the ones inside the case after removing the covers.
 Tighten or re-seal joints as necessary.
Inspect the appliance:
 Check the status of compressor aggregate – uniformity of operation, noise level, check the
performance of the compressor - measure time of pressurizing of air chamber from 0 to 6bar
(35 sec).
 Check the condition of the hangers above the pump.
 Checking the fan functioning.
 Examine the pump for:
­ impurities in the crankcase, free movement of the crank shaft, if necessary, replace any
faulty parts
18.5. Replacement of filters of air pump
Caution – hot surface! About disassembly the internal back cover can be hot the
internal surface.
 Disassembly the internal cover. These filter inserts are in
the top cover of crank case.
 Using screwdriver, remove the flexible guard.
 Remove the used and contaminated filters (A), (B).
 Insert a new set of input filter (B).
 Place prefilter (A) into the flexible guard (by plain side into
the crank case) and then to the opening.
18.6. Cleaning and disinfection of the exterior surfaces of the product
Use neutral agents for cleaning and disinfection of the external surfaces of the product.
The use of aggressive cleaning agents and disinfectants comprising alcohol and
chlorides may lead to the damage of the surface and the discoloration of the
product.
06/2020
- 27 -
NP-DK50 simple s DE-4_06-2020
Fig.8