Operating the field indicator
Menu
Operating
parameters
PARAM
Service
SERV
1)
Only, if DIM = TEXT
2)
Only, if NAMUR = Edit
3)
only available for service personnel
6.4
Configuration via interface & FieldCare Device Setup PC configu-
ration software
WARNING
L
The device is not explosion-protected when the housing is open.
The device must be configured outside the hazardous area.
The device can assume undefined states while configuring with FieldCare! This can result in
the undefined switching of outputs and relays.
To configure the device with the FieldCare Device Setup software, connect the device to your
PC. You need a special interface adapter for this purpose - the Commubox FXA291.
The 4-pin connector of the interface cable must be plugged into the corresponding socket
inside the device and the USB connector must be plugged into a free USB slot on the PC.
34
Downloaded from
www.Manualslib.com
Operating function
Parameter
Default /
Choice
User code
CODE
0 to 9 999
0
NAMUR
NAMUR
Standard
dEF
Edit
Edit
2)
NAMUR 20.5
N2050
NAMUR 3.80
20.5
to NAMUR
21.0
Service code
SCODE
----
manuals search engine
Operating function
Parameter
Default /
Choice
Program name
PNAME
2)
NAMUR 3.6
N_360
0 to NAMUR
3.60
20.5
2)
NAMUR 21.0
N2100
20.5 to
21.0
25 mA
3)
Parameter reset
PRSET
yes
YES
no
NO
RIA16
Operating function
Parameter
Default /
Choice
Firmware version
FWVER
2)
NAMUR 3.80
N_380
NAMUR 3.6
3.80
to NAMUR
20.5
Test
TEST
off
OFF
Open Collect.
OUT
Display
DISP
Endress+Hauser