ERROR CODE
E007.x
E009
E010
E011
E013
E014
E015
DISCRIPTION
PT calculation error.
The INR is too low.
The operation has timed out.
You inserted a used test strip,
or you applied a blood sample
to the test strip before the
meter warm-up was complete.
The strip is defective or im-
properly operated when adding
sample.
The value of HCT may be out
of range 30%-55%.
Bluetooth initialization error.
29
CORRECTIVE
ACTIONS
Retest with a new strip.
1. Make sure the strip code is
entered correctly.
2. Retest with a new strip.
Please ensure the sample is
added within 2 minute after
the 'add sample' indicator is
flashing.
Retest with a new strip. Make
sure to add sample within 2
minutes after the 'add blood'
indicator is flashing.
Retest using a new strip. Make
sure the operation of adding
sample is exactly correct
according to the user manual
guidance.
Patients with abnormal HCT
value may be recommended
to test the INR using another
method.
Turn off the meter and restart
the meter. If the problem still
exists, please contact the tech-
nician for further assistant.
EN