Cleaning and Disinfecting

Warnings:
- Cleaning and disinfecting are separate processes. Cleaning must precede disinfection. Before use
by a different individual: clean, disinfect, and check product for proper functioning.
- Follow all manufacturer bleach container and germicidal disinfectant safety guidelines.
- Note About Sterilization: High temperatures accelerate aging and will damage the product
assembly. The products featured in this manual are not packaged in a sterile condition, nor are
they intended or required to be sterilized prior to use. If institution protocol demands sterilization:
First, follow the cleaning and disinfecting instructions; then, open the inflation valve and use the
lowest possible sterilizing temperature, but not to exceed 70°C (158°F), for the shortest time
possible. DO NOT autoclave.
Cover Care
Note: Before cleaning and disinfecting, remove the cover from the cushion. Remove hook fastener pieces
from the bottom of the cover.
To Clean the Cover: Machine wash in cold water (30°C) with mild detergent, gentle
cycle, or wipe clean with neutral detergent and cold water. Tumble dry low.
To Disinfect the Cover: Hand wash in cold water (30°C) using 1 part
household liquid bleach per 9 parts water, then rinse thoroughly. Or
machine wash in hot water (60°C). Tumble dry low.
Cushion Care
Warning: Inflation valve(s) must be closed. DO NOT allow water or cleaning solution to enter the
cushion insert.
Cautions: DO NOT machine wash or machine dry the cushion insert or the contoured foam base.
Use of the following may cause damage to the cushion: abrasives (steel wool, scouring pads); caustic,
automatic dishwasher detergents; cleaning products that contain petroleum or organic solvents, including
acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover; oil-based
lotions, lanolin, or phenolic-based disinfectants; ozone gas; ultraviolet light cleaning methods.
To Clean the Cushion Insert: Carefully remove the cushion insert from the contoured
foam base. Close the inflation valve. Hand wash, using mild liquid hand soap, hand
dishwashing liquid, laundry detergent, or a multipurpose detergent (follow product
label instructions.) Gently scrub all cushion surfaces, using a soft plastic bristle
brush, a sponge or a cloth. Rinse with clean water. Air dry. DO NOT expose to direct
sunlight.
Caution: Rinse thoroughly. Detergent residue may cause air cells to stick together.
To Disinfect the Cushion Insert: Repeat the cleaning instructions, using 1
part household liquid bleach per 9 parts water. Keep the product wet with the
bleach solution for 10 minutes. Rinse thoroughly with clean water. Air dry.
Contoured Foam Base Care
To Clean the Contoured Foam Base: Remove the cover. Wipe gently with a cloth
dampened with water and a household detergent. Rinse cloth and re-wipe. Allow
foam to air dry thoroughly before putting the cover on.
Warning: Disinfectants are not effective on porous surfaces such as foam. If the foam
portion of the product becomes soiled or contaminated, DO NOT use with multiple individuals.
ROHO nexus SPIRIT Cushion Operation Manual
6
30C
30C
or
+
60C
30C
1:9
+
1:9
Wipe
with a
damp
cloth.