7.3

Operating conditions

Gauge pressure rated conditions
Installation conditions
Ambient conditions
Ambient temperature
Note
Measuring cell with silicone oil
filling
Display
Storage temperature
Climate class
Condensation
Degree of protection in
accordance with EN 60529
Electromagnetic compatibility
Interference emission and
interference immunity
Process medium conditions
Process temperature
Measuring cell with silicone oil
filling
With extension to Zone 0
Rated conditions for differential pressure and flow rate
Installation conditions
Installation instruction
Ambient conditions
Ambient temperature
Note
Measuring cell with silicone oil
filling
3000 kPa (435 psi)
measuring cell
Display
Storage temperature
Climate class
Condensation
Degree of protection in
accordance with EN 60529
Electromagnetic compatibility
Interference emission and
interference immunity
Process medium conditions
Process temperature
Measuring cell with silicone oil
filling
3000 kPa (435 psi)
measuring cell
36
Observe the temperature class in hazardous areas.
-40 ... +85 °C (-40 ... +185 °F)
-30 ... +85 °C (-22 ... +185 °F)
-50 ... +85 °C (-58 ... +185 °F)
Permitted
IP66, IP68
In accordance with EN 61326 and NAMUR NE 21
-40 ... +100 °C (-40 ... +212 °F)
-20 ... +60 °C (-4 ... +140 °F)
any
Observe the temperature class in hazardous areas.
-40 ... +85 °C (-40 ... +185 °F)
-20 ... +85 °C (-4 ... +185 °F)
For flow: -20 ... +85 °C (-4 ... +185 °F)
-30 ... +85 °C (-22 ... +185 °F)
-50 ... +85 °C (-58 ... +185 °F)
Permitted
IP66, IP68
In accordance with EN 61326 and NAMUR NE 21
-40 ... +100 °C (-40 ... +212 °F)
-20 ... +85 °C (-4 ... +185 °F)
SITRANS P310 with HART
A5E35075329-01, 05/2015