Connecting the HPLC Pump K–1001 to Other Instruments
WellChrom Solvent Organizer K–1500
Fig. 20
Fig. 21
SOP 20 Fixing the Solvent Organizer K-1500 above the Pump K-1001
For low pressure gradient mixing, the HPLC Pump K–1001 can be combined
with the KNAUER Solvent Organizer K–1500. In combination with the Solvent
Orga-nizer K–1500 the pump is able to deliver up to four different solvents in a
low pressure gradient (LPG) system. The HPLC Pump K–1001 controls all
electrical functions of the Solvent Organizer K–1500 and supplies electric
power.
HPLC Pump K–1001 combined with the Solvent Organizer K–1500
A
6.1
Pans for solvent bottles
6.2
LEDs A to D
6.3
Solvent inlet sites A to D
6.4
Solvent outlet sites A to D
Front View of the Solvent Organizer K–1500
The solvent organizer K-1500 can also be placed on top of the pump K-1001 or
on the pump together with the degasser. To avoid unintended moving of the
degasser, the organizer should be fixed using the 4 mounting brackets.
This SOP refers to the WellChrom HPLC Pump K-1001 and the Solvent
Organizer K–1500.
1. Remove the stainless steel pans (see Fig. 21, Pos 6.1) from the Solvent
Organizer K-1500.
2. Insert the 4 screws with washers through the slotts in the bottom of the
organizer and screw them into the boreholes of the brackets placed below
the organizer.
B
C
D
{6.1}
{6.2}
{6.3}
{6.4}