Operating the HPLC Pump K–1001
SOP 9
Fig. 14
SOP 10 Start and Stop of Solvent Delivery
SOP 11 Purging the HPLC Pump K–1001
2.
Type in the desired percentage of eluent A. The value in the right hand
adjacent field next to your input complements the value to give a total of
100%.
3.
Press any arrow key for confirmation.
Setting Events
Events are defined as states or changes of states of the electrical contacts of
the Events socket of the instrument. You can either open or close any of the
contacts or use a 500 ms puls. Some times the manual control may be usefull,
e.g. for test purposes. For connection of cables use sections „Electrical
Connections" on page 27
This SOP applies to the HPLC Pump K–1001 Firmware Revision 6.0X.
1.
Position the cursor at the event field.
2.
Start entry of events by pressing 0, 1, 2 or 3 for the first event. After every
entry, the underlined cursor jumps one position to the right. The following
table gives the meaning of the possible entries:
0 — Off
1 — On
2 — Pulse
3 — No change with respect to previous status
3.
Finish entry of events by pressing any arrow key.
Relations between event prgramming and event termial strip
This SOP applies to the HPLC Pump K–1001 Firmware Revision 6.0X.
1.
Apply SOP 7 „Setting the Flow Rate" on page 18.
2.
Use the START/STOP Key, see Pos. {2.1} in Fig. 2 on page 8, to start
and stop solvent delivery.
This SOP applies to the HPLC Pump K–1001 Firmware Revision 6.0X.
19