Herunterladen Inhalt Inhalt Diese Seite drucken

Sony UP-D55 Vor Verwendung Seite 16

Vorschau ausblenden Andere Handbücher für UP-D55:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 28
Recommended separation distances between portable and mobile RF communications equipment and the UP-D55
The UP-D55 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the UP-D55 can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (Transmitters) and the UP-D55 as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Caution
When you dispose of the unit or accessories, you must
obey the law in the relative area or country and the
regulation in the relative hospital.
Warning on power connection
Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) /
Appliance Connector / Plug with earthing-contacts
that conforms to the safety regulations of each
country if applicable.
2. Use the Power Cord (3-core mains lead) / Appliance
Connector / Plug conforming to the proper ratings
(Voltage, Ampere).
If you have questions on the use of the above Power
Cord / Appliance Connector / Plug, please consult a
qualified service personnel.
6
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
Warning on power connection for
medical use
Please use the following power supply cord.
With connectors (plug or female) and cord types other
than those indicated in this table, use the power supply
cord that is approved for use in your area.
Plug Type
Cord Type
Minimum Rating for Plug
and Appliance Couplers
Safety Approval
*Note: Grounding reliability can only be achieved when the equip-
ment is connected to an equivalent receptacle marked 'Hospital Only'
or 'Hospital Grade'.
For the customers in Europe
This product has been manufactured by or on behalf of
Sony Corporation, 1-7-1 Konan Minato-ku Tokyo, 108-
0075 Japan. Inquiries related to product compliance
based on European Union legislation shall be addressed
to the authorized representative, Sony Deutschland
GmbH, Hedelfinger Strasse 61, 70327 Stuttgart,
Germany.
For any service or guarantee matters, please refer to the
addresses provided in the separate service or guarantee
documents.
m
800 MHz to 2.5 GHz
d = 1.2 √P
d = 2.3 √P
0.12
0.23
0.38
0.73
1.2
3.8
12
United States and Canada
HOSPITAL GRADE*
Min.Type SJT
Min.18 AWG
10 A/125 V
UL Listed and CSA
2.3
7.3
23

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis