Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The GXS-700 sensor, used with a compliant computer, is intended for use in the electromagnetic
environment specified below. The customer or the user of the GXS-700 sensor should assure that it is
used in such an environment.
Immunity test
IEC 60601 test
Electrostatic
+ 6 kV contact
discharge
+ 8 kV air
(ESD)
IEC 61000-4-2
Electrical fast
+ 2 kV for
transient/burst
power supply
IEC 61000-4-4
+ 1 kV for
input/output lines
Surge
line(s) to line(s)
IEC 61000-4-5
line(s) to earth
Voltage dips,
<5% U
short interrup-
in U
T
tions and volt-
40% U
age variations
in U
T
on power sup-
70% U
ply input lines
in U
IEC 61000-4-
11
<5% U
in U
Power fre-
quency (50/60
Hz) magnetic
field
IEC 61000-4-8
NOTE 1 U
is the AC mains voltage prior to application of the test level.
T
(*) Computer used with the GXS-700 sensor must meet this rating.
B-2
Compliance level
level
Complies
Complies (*)
lines
+ 1 kV
Complies (*)
+ 2 kV
(95% dip
Complies (*)
T
) for 0,5 cycle
(60% dip
T
) for 5 cycles
(30% dip
T
) for 25 cy-
T
cles
(95% dip
T
) for 5 sec.
T
3A/m
Complies
Electromagnetic environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the GXS-700
requires continued operation during
power mains interruptions, it is
recommended that the GXS-700 be
powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.