PRODUCT MAINTENANCE

PRODUCT MAINTENANCE
16. PRODUCT MAINTENANCE
The operator should carry out device checks regularly in the intervals defined
by applicable regulations. Test results must be recorded.
Risk of servicing by persons without required qualification.
Repair works beyond standard maintenance (see chapter 16.1 - Maintenance
intervals) may only be performed by a qualified technician (organization
authorized by the manufacturer) or manufacturer's customer service.
Standard maintenance works (see Chapter 16.1 - Maintenance intervals) may
only be carried out by a trained operator.
Only use spare parts and accessories approved by the manufacturer.
The device has been designed and manufactured to keep its maintenance to a minimum. For
correct and reliable operation of the compressor perform the following operations.
Risk of injury and damage to the device.
Before any maintenance work, it is necessary to:
-
check, if it is possible to disconnect compressor from the supplied
equipment, so that there is no risk of damage to a person using the given
appliance or any other material damages
-
switch off the compressor
-
disconnect it from the mains (pull out the mains plug)
-
vent the compressed air from the air tank
Risk of injury when venting the compressed air.
Protect eyesight – wear goggles when venting the compressed air from the
compressor pneumatic system (air tank).
Risk of burn.
Pump components (head, cylinder, and pressure hose) have high temperature
during and shortly after compressor operation – do not touch these
components!
Let the device cool before any product maintenance, service or connection/
disconnection of pressurized air!
01/2020