Herunterladen Inhalt Inhalt Diese Seite drucken

BEHA UNITEST TELARIS 0100 Bedienungsanleitung Seite 52

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Technical Data
Measurement of trip Time tA:
Nominal fault current I∆N:
Type of test current:
Measurement time:
Measurement Range:
Resolution:
Tolerance:
Current-cut off:
Measurement of Ground Resistance:
Measurement current:
Measurement time:
Measurement Range:
Resolution:
Tolerance:
Current-cut off:
Measurement of trip Current I
Measurement Current:
Measurement time:
Resolution:
Tolerance:
Current cut off:
24 month Warranty
UNITEST instruments are subject to strict quality control. However, should the instrument function improperly
during daily use, you are protected by our 24 months warranty (valid only with invoice).
We will repair free of charge any defects in workmanship or material, provided the instrument is returned un-
opened and untampered with, i.e. with undamaged warranty label.
Any damage due to dropping or incorrect handling are not covered by the warranty.
If the instrument shows failure following expiration of warranty, our service department can offer you a quick and
economical repair.
Subject to technical changes without notice!
52
10, 30, 100, 300, 500mA standard
100, 300, 500 mA time delayed
AC
500ms , 300ms when measuring trip current
1...500ms
1ms
± (2% rdg. +2Digit)
immediately for RCD tripping or when exeeding
contact voltage (25V / 50V)
1/3 nominal fault current I∆N
200ms each. immediate interruption when exceeding threshold values
10...1999Ω , 0,20kΩ ... 19,99kΩ
1 Ω, 0,01kΩ
±6% rdg.. ±10Digit / ±6% rdg. ±20Digit at I∆N = 10mA
immediately for RCD tripping or when exeeding
contact voltage (25V / 50V)
:
40 ... 140% of the nominal fault current I∆N
200ms each. immediate interruption when exceeding threshold values
10% I∆N
-0% ... +10%
immediately for RCD tripping or when exeeding
contact voltage (25V / 50V)

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis