CALIBRATION MODE

The TB82EC (four-electrode conductivity) and TB82TC (Toroidal Conductivity) transmitters often require a wet calibration
of the sensor for greatest accuracy. Two-electrode sensors, used with the TB82TE transmitter, do not require wet
calibration if the cell constant has been properly programmed into the analyzer.
calibration of sensors when in the process or in standard solutions. The CALIBRATION mode also allows calibration of
the temperature input and fine tuning the 4-20 mA output for non-fieldbus versions.

Calibration Flow Chart for TB82EC (Four Electrode) and TB82TE (Two Electrode)

MEASURE
CALIBRATE

Calibration Flow Chart for TB82TC (Toroidal Conductivity)

MEASURE
CALIBRATE
Caution.
The Reset Calibrate State will reset all calibration
values; therefore, the process sensor and
temperature sensor will require calibration after
performing the Reset Calibrate procedure.
TB82COND-ATEX-EN-A
CON Cal
Stable?
Temperature Cal
Enter Temp
Edit Calibration
Enter Slope
Enter Offset
Reset Cal
Output Cal
Adjust 4 mA
Adjust 20 mA
CON Cal
Zero Point
Span Point
Temperature Cal
Enter Temp
Edit Calibration
Enter Slope
Enter Offset
Reset Cal
Output Cal
Adjust 4 mA
Adjust 20 mA
TB82 Conductivity Series ATEX Product Instruction Manual
Enter New Value
Enter Zero Value
Stable? Yes/No
Enter Span Level
Stable? Yes/No
The CALIBRATION mode allows